Biological Incubator
Specifications
- Origin
- China
- Price Term
- FOB
Biological Incubator is a Biological Chemical Products product. Key specs — Origin: China; Price Term: FOB.
Product Description
I. Description:
Biochemical incubator with two-way thermostat cooling and heating systems, temperature control features, plants, biology, microbiology, genetics, viruses, medicine, environmental protection, scientific research and teaching departments indispensable laboratory equipment, widely used in thermostats experiment, The culture experiment, the environment, and other experiments. Its main features: This product is a new type of intelligent temperature control for temperature control, LED figures show that the visibility; Imported stainless steel interior, strong and corrosion resistance, and other advantages.
II. Specifications:
1. Power: 220 V 50HZ
2. Refrigeration Power: 105 W
3. Heating Power: 100 W
4. With lighting
5. Temperature range: 5-50 degree (digital LED, with observation windows)
6. Temperature accuracy: 0.5 degree
7. Effective Capacity: 150 L
8. Time: Continuous
III. Instructions for use:
1. Incubators should be placed dry, clean, well-ventilated and away from sources of heat and the sun and indoor work, placing a smooth, to prevent vibration noise.
2. Before use, the control switch on the panel are in a non-working state.
3. Placed in the incubator shelves experimental samples, placed at various experimental bottle (or containers) should maintain a proper distance, so cold (hot) air convection cycle.
4. Connected to power, the power switch will be placed in "open" (ON) position, bright lights, LED digital display that shows the actual temperature inside.
5. Based on the use of smart temperature controller that you set the temperature, the temperature controller will enable automatic control box to maintain a constant temperature (when the box when the temperature is higher than setting incubator will automatically refrigeration, but training Box will automatically heating).
IV. Maintenance and Warranty:
1. Incubators should be reliable grounding.
2. To ensure the effective condensation heat, the distance between the condenser and walls more than 100 mm. Box side should be 50 mm gap, the top box space at least 30 mm.
3. Incubators in the removal, repair, maintenance, keen to avoid the collision, the largest gradient of less than 45 degrees.
4. Remove the outlet, have five minutes to insert, such as immediately inserted, compressor components easily damaged.
5. The incubator refrigeration work, it is not appropriate to the ambient temperature inside temperature and the difference is greater than 25 degrees.
6. Containing prohibited chemical-volatile solvents, gases and explosive placed inside a flammable gas, the incubator can not be used near flammable aerosols, so as to avoid electric spark ignition.
7. To prevent pollution, the use of low-temperature should be avoided as far as possible in the work of interior wall condensation Bingshuang.
8. Refrigeration system work, should avoid frequent opening of the door, to keep the temperature stability and to prevent dust, dirt access would benefit.
9. Refrigeration system stopped working, with a soft cloth Cajing studio.